Comparative evaluation of immunochromatographic test and ELISA for detection of HBsAg and anti-HCV among patients at a tertiary care hospital in South Kashmir.
- Abiroo Jan , Associate Professor (Deptt of Microbiology ,GMC Anantnag).
- Aamina Samreen , Senior Resident (Department of Microbiology, GMC Anantnag).
- Mir Haris , Scientist-B (VRDL GMC Anantnag).
- Iqra Hussain Shah , Scientist B Medical (VRDL GMC Anantnag).
- Nadeem Ahmad Bhat, , Research Assistant (VRDL GMC Anantnag).
- Jahangir Mohiuddin , Laboratory Technician (VRDL GMC Anantnag).
- Irfan Rafiq Dar ,
- Tufail Ahmed Wani. ,
Article Information:
Abstract:
Background: Hepatitis B virus (HBV) and hepatitis C virus (HCV) infections remain significant public health concerns in Kashmir. Rapid immunochromatographic tests (ICTs) allow decentralized screening, but their accuracy compared with enzyme-linked immunosorbent assay (ELISA) requires local validation. Aim: To evaluate the diagnostic performance of commonly used rapid ICTs against ELISA for detecting hepatitis B surface antigen (HBsAg) and anti-HCV antibodies in patients attending a tertiary care hospital in South Kashmir. Materials and Methods: This retrospective cross-sectional diagnostic accuracy study included 3,951 paired serum samples tested by ICT and ELISA between January 2024 and June 2025. ELISA was the reference standard. Sensitivity, specificity, predictive values, likelihood ratios, Cohen’s kappa, and McNemar’s exact test were calculated with Wilson 95% confidence intervals. Subgroup analyses were conducted by sex and age (<20 vs. ≥20 years). Results: ELISA identified HBsAg prevalence of 5.7% (95% CI: 5.0–6.5) and anti-HCV prevalence of 11.0% (95% CI: 10.1–12.1). For HBsAg, ICT showed 89.3% sensitivity and 99.95% specificity (LR+ ≈ 1,664; LR− = 0.11), with strong agreement (κ ≈ 0.936). McNemar’s test indicated significantly more false negatives than false positives (p ≈ 1×10⁻⁵). For anti-HCV, ICT sensitivity was 94.7% and specificity 99.4% (LR+ ≈ 158.6; LR− ≈ 0.05), with strong agreement (κ ≈ 0.943) and no significant discordance (p ≈ 0.88). No significant differences were observed by sex or age. Conclusion: Rapid ICTs demonstrated excellent specificity and high sensitivity for hepatitis screening. However, excess false-negative HBsAg results support reflex ELISA testing of ICT-negative samples, particularly in high-risk groups, to reduce missed diagnoses.
Keywords:
Article :
INTRODUCTION:
Chronic infections with hepatitis B virus (HBV) and hepatitis C virus (HCV) are leading causes of liver-related morbidity and mortality worldwide, contributing substantially to cirrhosis and hepatocellular carcinoma [1,2]. Globally, an estimated 250 million people live with chronic HBV infection and approximately 71 million with chronic HCV infection, with South Asia bearing a disproportionate share of this burden [3]. In India, and particularly in the Kashmir region, transmission is driven by injection drug use, unsafe medical injections, and inadequate infection control practices. Reported regional seroprevalence estimates range from 1–5% for HBsAg and 2–10% for anti-HCV [4,5].
Serological detection remains the cornerstone of hepatitis screening. While ELISA-based assays offer high sensitivity and reliability, they require laboratory infrastructure, trained personnel, and batch processing, which may delay diagnosis in resource-limited or high-throughput settings [6]. In contrast, immunochromatographic tests (ICTs) provide rapid, point-of-care results and are widely deployed under India’s National Viral Hepatitis Control Program [7]. However, the diagnostic performance of ICTs varies across manufacturers and settings, with reported sensitivities ranging from 50–95% for HBsAg and 70–98% for anti-HCV [8–10]. False-negative results, in particular, pose a risk of delayed diagnosis and ongoing transmission.
Given the rising burden of viral hepatitis among young adults in Kashmir and the extensive reliance on ICTs for screening, local validation of test performance is essential [4,5]. This study aimed to evaluate the diagnostic accuracy of commonly used ICTs against ELISA for detection of HBsAg and anti-HCV in a large cohort from a tertiary care hospital in South Kashmir, with additional assessment of agreement metrics and subgroup variations by sex and age.
MATERIALS AND METHODS:
Study Design and Setting
A retrospective, cross-sectional diagnostic accuracy study was conducted at the Department of Microbiology, Government Medical College Hospital, Anantnag, South Kashmir. The study adhered to the STARD 2015 reporting guidelines for diagnostic accuracy studies [11].
Study Population
Residual serum samples from patients tested for HBsAg and/or anti-HCV between 1 January 2024 and 30 June 2025 were eligible. Inclusion criteria comprised availability of paired ICT and ELISA results for the same sample. Samples with incomplete test pairs, hemolysis, or non-human origin were excluded. Consecutive non-probability sampling of all eligible samples yielded a final sample size of 3,951.
Ethical Considerations
The study protocol was approved by the Institutional Ethics Committee of Government Medical College, Anantnag (IEC/GMC/2024/012). As the analysis was retrospective and based on de-identified laboratory records, informed consent was waived in accordance with Indian Council of Medical Research guidelines.
Description of Diagnostic Tests
The index tests were commercially available immunochromatographic card tests for HBsAg and anti-HCV (e.g., SD Bioline, Alere), performed according to manufacturer instructions by trained laboratory personnel. Results were read visually at 15–20 minutes. The reference standard comprised third- or fourth-generation ELISA kits (e.g., Erba Lisa, Transasia) processed in batches with appropriate positive and negative controls. ICT and ELISA results were generated as part of routine workflow, with no post-hoc reinterpretation.
Data Collection and Statistical Analysis
De-identified demographic and laboratory data (age, sex, ICT result, ELISA result) were extracted from laboratory registers into a spreadsheet and analyzed using SPSS version 20 (IBM Corp., Armonk, NY, USA). Continuous variables were summarized as medians with interquartile ranges (IQR), and categorical variables as frequencies and percentages. Diagnostic indices were derived from 2×2 contingency tables. Sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were calculated using standard formulas, with Wilson 95% confidence intervals [12]. Positive and negative likelihood ratios (LR+ and LR−) were calculated using exact cell counts; likelihood ratio confidence intervals were estimated using the method of Simel et al. Agreement between ICT and ELISA was assessed using Cohen’s kappa with 95% confidence intervals [13]. Discordance was evaluated using McNemar’s exact binomial test, given the small number of discordant pairs [14]. Subgroup analyses by sex and age (<20 vs. ≥20 years) were performed using chi-square tests. A two-sided p-value <0.05 was considered statistically significant.
RESULTS:
Cohort Characteristics
A total of 3,951 paired serum samples were analyzed. Of these, 1,996 (50.5%) were from male patients and 1,955 (49.5%) from female patients. The median age was 23 years (IQR: 19–28). By ELISA, 225 samples (5.7%) were positive for HBsAg and 436 (11.0%) were positive for anti-HCV (Table 1).
Table 1: Cohort Characteristics (N = 3951)
|
Total Patients |
Male |
Female |
ELISA Positive HBsAg |
ELISA Positive Anti HCV |
|
3951 |
1996 (50.5%) |
1955 (49.5%) |
225 (5.7%) |
436 (11.0%) |
Diagnostic Performance for HBsAg
Among 225 ELISA-confirmed HBsAg-positive samples, ICT correctly identified 201 true positives and missed 24 cases, yielding a sensitivity of 89.3% (95% CI: 84.7–92.8). Among 3,726 ELISA-negative samples, ICT correctly classified 3,724 as negative, with two false-positive results, corresponding to a specificity of 99.95% (95% CI: 99.8–100.0). The PPV and NPV were 99.0% and 99.4%, respectively. The LR+ was approximately 1,664.3, reflecting near-perfect specificity, while the LR− was 0.11. Agreement between ICT and ELISA was strong (Cohen’s κ ≈ 0.936). McNemar’s exact test demonstrated a statistically significant excess of false-negative over false-positive ICT results (FP = 2 vs. FN = 24; two-sided p ≈ 1×10⁻⁵), indicating systematic under-detection of HBsAg by ICT (Table 2).
Table 2. Diagnostic Performance of ICT for HBsAg (Reference: ELISA)
|
ICT Result |
ELISA Positive (225) |
ELISA Negative (3726) |
|
Positive |
201 |
2 |
|
Negative |
24 |
3724 |
Sensitivity: 89.3% (95%CI: 84.7–92.8)
Specificity: 99.95% (95%CI: 99.8–100.0)
PPV: 99.0%
NPV: 99.4%
LR+: ≈1,664.3; LR−: ≈0.11
Cohen’s κ: ≈ 0.936
McNemar’s exact p-value: ≈ 1×10⁻⁵
Diagnostic Performance for Anti-HCV
For anti-HCV, ICT detected 413 true positives and 23 false negatives among 436 ELISA-positive samples, giving a sensitivity of 94.7% (95% CI: 92.1–96.5). Of 3,515 ELISA-negative samples, 3,494 were correctly identified as negative, with 21 false positives, resulting in a specificity of 99.4% (95% CI: 99.0–99.6). The PPV was 95.2% and the NPV was 99.3%. The LR+ was approximately 158.6 and the LR− was 0.05. Agreement was strong (κ ≈ 0.943). McNemar’s exact test showed no significant asymmetry between false-positive and false-negative results (FP = 21 vs. FN = 23; two-sided p ≈ 0.88) (Table 3).
Table 3. Diagnostic Performance of ICT for Anti-HCV (Reference: ELISA)
|
ICT Result |
ELISA Positive (436) |
ELISA Negative (3515) |
|
Positive |
413 |
21 |
|
Negative |
23 |
3494 |
Sensitivity: 94.7% (95%CI: 92.1–96.5)
Specificity: 99.4% (95%CI: 99.0–99.6)
PPV: 95.2%
NPV: 99.3%
LR+: ≈158.6; LR−: ≈0.05
Cohen’s κ: ≈0.943
McNemar’s exact p-value: ≈0.88
Subgroup Analyses
HBsAg sensitivity was slightly higher in females (91.8%) than males (87.4%) and in individuals aged ≥20 years (90.6%) compared with those <20 years (86.8%), but these differences were not statistically significant (p = 0.21). Anti-HCV sensitivity showed minimal variation by sex or age, with no significant subgroup differences (Table 4).
Table 4. Shows the Sensitivity of ELISA in terms of gender
|
Marker |
sex |
Sensitivity % (95% CI) |
ELISApositive |
|
HbsAg |
Male |
87.4 (80.5–92.5) |
127 |
|
Female |
91.8 (84.5–96.2) |
98 |
|
|
Anti-HCV |
Male |
95.2 (91.7–97.4) |
249 |
|
Female |
94.1 (89.7–96.9) |
187 |
Table 5.Shows the Sensitivity of ELISA in terms of Age Group.
|
Marker |
Age Group |
Sensitivity % (95% CI) |
ELISA positive |
|
HbsAg
|
<20 years |
86.8 (77.1–93.2) |
76 |
|
≥20 years |
90.6 (84.8–94.6) |
149 |
|
|
Anti-HCV |
<20 years |
95.7 (90.3–98.5) |
117 |
|
≥20 years |
94.4 (91.2–96.6) |
319 |
DISCUSSION:
In this large hospital-based evaluation from South Kashmir, rapid ICTs demonstrated excellent specificity (>99%) and high sensitivity for detection of both HBsAg and anti-HCV when compared with ELISA. The sensitivity for HBsAg (89.3%) was within the range reported in prior Indian and international studies, while anti-HCV sensitivity was higher (94.7%), consistent with recent meta-analytic estimates [8,10,17].
A key finding was the statistically significant excess of false-negative ICT results for HBsAg, as demonstrated by McNemar’s exact test. This asymmetry indicates that while false positives are rare, a meaningful proportion of true HBsAg-positive cases may be missed if ICT is used as the sole screening tool. In contrast, no such systematic discordance was observed for anti-HCV. These results reinforce concerns that ICT-based screening, particularly for HBV, may under-detect low-titer or early infections, potentially delaying diagnosis and facilitating ongoing transmission in high-risk populations [4,5,18].
Subgroup analyses suggested modest variations in sensitivity by sex and age, with slightly lower HBsAg sensitivity among males and individuals under 20 years of age, although these differences were not statistically significant. Such patterns may reflect biological or behavioral factors and warrant further investigation in larger, prospective studies.
Compared with published benchmarks from blood donor and hospital-based studies in India, the specificities observed here exceed 95%, supporting the continued use of ICTs for initial triage under national screening programs [7,23–25]. However, the demonstrated false-negative burden for HBsAg underscores the need for reflex ELISA testing of ICT-negative samples in clinically suspected or high-risk cases. Strengths of this study include its large sample size and comprehensive evaluation of agreement and discordance metrics. Limitations include its retrospective design, single-center setting, and lack of nucleic acid testing for confirmatory diagnosis.
CONCLUSION:
Rapid immunochromatographic tests show excellent specificity and good sensitivity for hepatitis screening in South Kashmir. However, the significant excess of false-negative ICT results for HBsAg highlights the need for reflex ELISA testing, particularly in high-risk or clinically suspected cases, to minimize missed diagnoses and support hepatitis elimination goals.
REFERENCES:
1. World Health Organization. Global hepatitis report 2023. Geneva: WHO; 2023.
2. Thomas DL. Global elimination of chronic hepatitis. New England Journal of Medicine. 2019 May 23;380(21):204150.
3. Lavanchy D. Evolving epidemiology of hepatitis C virus. Clinical microbiology and infection. 2011 Feb 1;17(2):10715.
4. Khuroo SR, Nazir H, Talwar Y, Singh JP, Chowdhary P, Kaur R, Khan UR, Manzoor A, SINGH Sr JP, Khan U, Manzoor Sr A. Risk Factors Contributing to Hepatitis C Virus Infection in Kashmir: A Meta-Analysis. Cureus. 2025 Jun 5;17(6).
5. Wani KF, Rather YH, Kousar MU, Bhat FR, Rasool U. Seroprevalence of HIV, HCV, and HBV: The Associated Psychiatric Co-morbidities Among Injection Drug Users in Kashmir—A Hospital-Based Cross-sectional Study. Indian Journal of Psychological Medicine. 2024 Jun 16:02537176241257673.
6. Gish RG, Locarnini S, Thio CL, et al. Point-of-care tests for hepatitis B: an overview. Cells 2020;9:2233.
7. Kumar A, Acharya SK, Singh SP, Arora A, Dhiman RK, Aggarwal R, Anand AC, Bhangui P, Chawla YK, Gupta SD, Dixit VK. 2019 update of Indian national association for study of the liver consensus on prevention, diagnosis, and management of hepatocellular carcinoma in India: the Puri II recommendations. Journal of clinical and experimental hepatology. 2020 Jan 1;10(1):43-80.
8. Tang W, Chen W, Amini A, Boeras D, Falconer J, Kelly H, Peeling R, Varsaneux O, Tucker JD, Easterbrook P. Diagnostic accuracy of tests to detect Hepatitis C antibody: a meta-analysis and review of the literature. BMC infectious diseases. 2017 Nov 1;17(Suppl 1):695.
9. Njouom R, Ndjoumou AR, Ndalla P, et al. Evaluation of HBsAg rapid test performance for different biological samples in Burkina Faso. BMC Infect Dis 2015;15:548.
10. Goyal A, Finlayson AE, Leelavathi L, et al. Diagnostic accuracy of rapid diagnostic tests for hepatitis B virus surface antigen: a systematic review and meta-analysis. J Clin Microbiol 2020;58:e01798-19.
11. Bossuyt PM, Reitsma JB, Bruns DE, Gatsonis CA, Glasziou PP, Irwig L, Lijmer JG, Moher D, Rennie D, De Vet HC, Kressel HY. STARD 2015: an updated list of essential items for reporting diagnostic accuracy studies. Radiology. 2015 Dec;277(3):826-32.
12. Newcombe RG. Two‐sided confidence intervals for the single proportion: comparison of seven methods. Statistics in medicine. 1998 Apr 30;17(8):857-72.
13. Cohen J. A coefficient of agreement for nominal scales. Educ Psychol Meas 1960;20:37-46.
14. McNemar Q. Note on the sampling error of the difference between correlated proportions or percentages. Psychometrika 1947;12:153-7.
15. Nazir M, Yousuf R, Amin M, et al. A comparative study of screening of hepatitis B by two different immunochromatographic methods among patients attending a tertiary care hospital. Int J Curr Microbiol App Sci 2019;8:1506-13.
16. Ghosh S, Mandal S, Chakrabarti S. Comparison of ELISA and ICT for detection of hepatitis viruses in blood donors. Asian J Transfus Sci 2021;15:10-4.
17. Moses SE, Sudhakar R, Suresh P, et al. Diagnostic accuracy of rapid tests for detection of antibodies to hepatitis C virus in India: a systematic review and meta-analysis. J Fam Med Prim Care 2019;8:2723-9.
18. Heberling RL, Kalter SS. Rapid dot-immunobinding assay on nitrocellulose for viral antibodies. Journal of clinical microbiology. 1986 Jan;23(1):109-13.
19. Allain JP, Candotti D, Soldan K, Sarkodie F, Phelps B, Giachetti C, Shyamala V, Yeboah F, Anokwa M, Owusu-Ofori S, Opare-Sem O. The risk of hepatitis B virus infection by transfusion in Kumasi, Ghana. Blood, The Journal of the American Society of Hematology. 2003 Mar 15;101(6):2419-25.
20. Mandawat BK, Mathur V. A comparative evaluation for detection of hepatitis B surface antigen and anti-HCV antibodies by rapid diagnostic tests and ELISA methods. Int J Curr Microbiol App Sci 2019;8:1959-64.
21. Botheju WSP, Zghyer F, Mahmud S, et al. Global meta-analysis on the association between diagnostic test accuracy and hepatitis C virus prevalence. Liver Int 2023;43:219-29.
22. Singh KP, Sehgal V, Gupta S. Performance of rapid immuno-chromatographic test in comparison with ELISA for detection of anti-HCV antibodies among blood donors. J Fam Med Prim Care 2020;9:3035-9.
23. Kumar R, Kaur A, Gupta S, et al. Comparison of rapid immunochromatographic card tests with ELISA for diagnosis of hepatitis B virus infection in patients attending tertiary care centre in Punjab, India. Int J Res Med Sci 2018;6:341-5.
24. Saini N, Singh S, Gupta V. Evaluation of screening effectiveness of hepatitis B surface antigen and anti-HCV rapid test kits in a tertiary care hospital in North India. J Clin Diagn Res 2021;15:DC15-9.
25. Al-Mosawy AJ, Al-Hamami NA, Al-Khafaji ZM. Comparison of rapid anti-HCV multi-sure kit with gold standard ELISA in blood donors. J Clin Diagn Res 2020;14:DC01-4.
26. Awan UA, Waqar W, Chaudhary MA, et al. Comparison of different ICT kits for HBsAg and anti-HCV using gold standard ELISA. Pak J Med Res 2012;51:72-6.